Seven commercially available biliary stents were tested to determine i
f they caused any clinically important attenuation of therapeutic radi
ation. Attenuation effects were evaluated in models of brachytherapy (
iridium-192 source) and external beam radiation therapy (cobalt-60 sou
rce), with use of parameters that mimicked those used in patient treat
ment for biliary neoplasms. No stent demonstrated clinically important
attenuation or scatter of therapeutic radiation, so local dosimetry w
as not affected.