DIFFERENTIAL INTERFERENCE CONTRAST MICROSCOPE WITH DIFFERENTIAL DETECTION FOR OPTIMIZING IMAGE-CONTRAST

Citation
H. Ooki et al., DIFFERENTIAL INTERFERENCE CONTRAST MICROSCOPE WITH DIFFERENTIAL DETECTION FOR OPTIMIZING IMAGE-CONTRAST, Applied optics, 35(13), 1996, pp. 2230-2234
Citations number
7
Categorie Soggetti
Optics
Journal title
ISSN journal
00036935
Volume
35
Issue
13
Year of publication
1996
Pages
2230 - 2234
Database
ISI
SICI code
0003-6935(1996)35:13<2230:DICMWD>2.0.ZU;2-Y
Abstract
A laser scanning differential interference contrast microscope using a differential detection method is proposed. This microscope permits ob servers to control the differential image contrast with a simple opera tion. Utilizing this function, the observers are capable of obtaining the-optimized differential image whose contrast is most favorable for observation. A simple theoretical analysis and experimental considerat ions are described. (C) 1996 Optical Society of America