STUDY OF A VERY LONG SCINTILLATING-FIBER TOF DETECTOR - APPLICATION TO THE TAU-CHARM-FACTORY

Citation
A. Ferrer et al., STUDY OF A VERY LONG SCINTILLATING-FIBER TOF DETECTOR - APPLICATION TO THE TAU-CHARM-FACTORY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 371(3), 1996, pp. 397-405
Citations number
14
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
371
Issue
3
Year of publication
1996
Pages
397 - 405
Database
ISI
SICI code
0168-9002(1996)371:3<397:SOAVLS>2.0.ZU;2-E
Abstract
We have studied the time resolution, light yield and attenuation lengt h of a scintillation counter prototype made of a bundle of plastic sci ntillating fibers 2 mm in diameter 6.15 m long and 5 cm X 5.5 cm in cr oss section. In order to achieve 3 sigma K-pi separation at 1 GeV/c vi a TOF in the Tau-Charm-Factory geometry one requires approximate to 12 0 ps time resolution. Measurements using one photomultiplier on cosmic muons crossing through the center of the counter give a resolution si gma = (237+/-5) ps thus indicating that two layers of counters each lo oked at by photomultipliers at both ends can provide the needed time r esolution.