A 3-DETECTOR POSITRON LIFETIME SETUP SUITED FOR MEASUREMENTS ON IRRADIATED STEELS

Citation
L. Vanhoorebeke et al., A 3-DETECTOR POSITRON LIFETIME SETUP SUITED FOR MEASUREMENTS ON IRRADIATED STEELS, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 371(3), 1996, pp. 566-571
Citations number
11
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
371
Issue
3
Year of publication
1996
Pages
566 - 571
Database
ISI
SICI code
0168-9002(1996)371:3<566:A3PLSS>2.0.ZU;2-0
Abstract
An advanced positron lifetime setup is described, particularly suited to perform measurements on irradiated nuclear reactor pressure vessel steel. The equipment exists of three BaF2-detectors operated in triple coincidence mode to avoid the very disturbing Co-60-contribution in t he lifetime spectra. The use of BaF2 scintillators gives rise to an ac ceptable efficiency and a time resolution well below 300 ps.