2-STAGE PROCEDURE FOR EVALUATING INTERASSAY CARRYOVER ON RANDOM-ACCESS INSTRUMENTS

Citation
B. Schlain et al., 2-STAGE PROCEDURE FOR EVALUATING INTERASSAY CARRYOVER ON RANDOM-ACCESS INSTRUMENTS, Clinical chemistry, 42(5), 1996, pp. 725-731
Citations number
14
Categorie Soggetti
Chemistry Medicinal
Journal title
ISSN journal
00099147
Volume
42
Issue
5
Year of publication
1996
Pages
725 - 731
Database
ISI
SICI code
0009-9147(1996)42:5<725:2PFEIC>2.0.ZU;2-7
Abstract
A two-stage statistical procedure based on Dunnett's multiple comparis on procedures with a control has been developed for detecting interass ay carryover biases on the Abbott AxSYM(TM) System, a random- and cont inuous-access immunoanalyzer. With this procedure, every potential sou rce of interassay carryover can be tested and estimated, In minimizing required sample sizes, the first stage is used primarily to detect an d eliminate from further testing the assay reagent sources that do not cause carryover biases and the assay sources that cause very large ca rryover biases. Retested more extensively in the second testing stage are the cases where the data from the first testing stage are insuffic ient for judgment. An example data set from the Abbott AxSYM Free T-4 assay is used to illustrate the methodology.