PERFORMANCE OF FRAGMENT SEPARATOR WITH HOMOGENEOUSLY THICK ENERGY DEGRADER - A HIGHLY FLEXIBLE SYSTEM FOR USE WITH DEGRADERS IN A WIDE THICKNESS RANGE

Citation
S. Mitsuoka et al., PERFORMANCE OF FRAGMENT SEPARATOR WITH HOMOGENEOUSLY THICK ENERGY DEGRADER - A HIGHLY FLEXIBLE SYSTEM FOR USE WITH DEGRADERS IN A WIDE THICKNESS RANGE, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 372(3), 1996, pp. 489-500
Citations number
13
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
372
Issue
3
Year of publication
1996
Pages
489 - 500
Database
ISI
SICI code
0168-9002(1996)372:3<489:POFSWH>2.0.ZU;2-5
Abstract
A projectile-fragment separator, secondary beam line, has been constru cted at RCNP for use with the K = 400 ring cyclotron. Products of the projectile fragmentation reactions are isotopically separated by makin g use of their energy-loss effect in an energy degrader. Unlike the ex isting separators of similar kind, a homogeneously thick degrader is u sed and achromatic focus is achieved by the optics of the magnetic sys tem. The separator was designed so as to possess a high ion-optical fl exibility which allows the use of degraders of widely ranged thickness . For commissioning the separator, a detailed experimental analysis of the optics was carried our to realize the design performance, and a p rescription to correctly determine the strengths of focusing devices w as established. The operational performance of the separator is presen ted.