A DUAL VACUUM CHAMBER FOURIER-TRANSFORM MASS-SPECTROMETER WITH RAPIDLY INTERCHANGEABLE LSIMS, MALDI, AND ESI SOURCES - INITIAL RESULTS WITHLSIMS AND MALDI
Ja. Carroll et al., A DUAL VACUUM CHAMBER FOURIER-TRANSFORM MASS-SPECTROMETER WITH RAPIDLY INTERCHANGEABLE LSIMS, MALDI, AND ESI SOURCES - INITIAL RESULTS WITHLSIMS AND MALDI, Analytical chemistry, 68(10), 1996, pp. 1798-1804
A design is presented involving two separate vacuum chambers to provid
e nearly simultaneous capabilities of liquid secondary ion mass spectr
ometry (LSIMS), matrix-assisted laser desorption/ionization (MALDI), a
nd electrospray ionization (ESI) in an external source Fourier transfo
rm mass spectrometer. The instrument consists of two vacuum chambers,
one with five stages of differential pumping for ESI and another with
two stages of differential pumping for a combined LSIMS/MALDI source,
The chamber dedicated to ESI was formerly a three-stage chamber with L
SIMS and electron ionization. Two additional stages were added with th
e ESI source, LSIMS and MALDI have similar vacuum requirements and wer
e moved to a newly built chamber with two stages of pumping. We presen
t our first results obtained on the new vacuum chamber, Data presented
for the MALDI source show that, with only two stages of pumping, and
with shorter radio frequency-only quadrupole rods for ion injection, s
pectra comparable to those obtained on the formerly three-stage instru
ment can be obtained, Characterization of the MALDI source and data on
linear, cyclic, and branched oligosaccharides are given, Finally, the
design of the second chamber is proposed as a low-cost prototype for
an external source FTMS instrument.