INTERNATIONAL COMPARISON OF STABILIZED HE-NE LASERS BY THE SATURATED ABSORPTION OF I-127(2) AT LAMBDA-APPROXIMATE-TO-633 NM INVOLVING THE CEM (SPAIN), THE IPQ (PORTUGAL) AND THE BIPM

Citation
E. Prieto et al., INTERNATIONAL COMPARISON OF STABILIZED HE-NE LASERS BY THE SATURATED ABSORPTION OF I-127(2) AT LAMBDA-APPROXIMATE-TO-633 NM INVOLVING THE CEM (SPAIN), THE IPQ (PORTUGAL) AND THE BIPM, Metrologia, 32(5), 1996, pp. 379-384
Citations number
17
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
Journal title
ISSN journal
00261394
Volume
32
Issue
5
Year of publication
1996
Pages
379 - 384
Database
ISI
SICI code
0026-1394(1996)32:5<379:ICOSHL>2.0.ZU;2-W
Abstract
The results of an international comparison of iodine-stabilized He-Ne lasers at lambda approximate to 633 nm, made by beat-frequency techniq ues, between the CEM (Spain), the IPQ (Portugal) and the BIPM are pres ented. Four lasers took part in the ten-day period of measurements, an d were compared once or twice a day. The frequency differences between the lasers were all less than 6 kHz and the highest standard uncertai nty of the mean of the frequency differences between two lasers was 7 kHz. These results lie inside the standard uncertainty given in the mi se en pratique of the definition of the metre. Parameters which affect the frequency of the lasers, such as modulation width, iodine pressur e and laser power, were also investigated.