Aa. Nikonov, A LOW-TEMPERATURE DEVICE FOR MEASURING DIFFERENTIAL MAGNETIC-SUSCEPTIBILITY VIA DOUBLE LOCK-IN DETECTION, Instruments and experimental techniques, 38(6), 1995, pp. 807-809
This paper describes a device for measuring differential magnetic susc
eptibility chi = partial derivative M/partial derivative H in a temper
ature range of 4.2 to 350 degrees K under a low ac magnetic field of 0
.01-30 Oe oscillating at a frequency of 1-20 kHz. The sensitivity thre
shold of the device can be reduced to 10(-9) CGSM unit (10(-10) A . m(
2)) in a ac field of 3 Oe.