Wedge factor (WF) variation with field size and depth has been analyse
d for 6 and 15 MV X-ray beams. The measured field size effect is 1.4%
in the worst case, whereas the WF shows a linear dependence with depth
giving a maximum variation of 0.22%/cm (60 degrees wedge, 6 MV beam).
An empirical correction for WF depth dependence on wedge thickness is
proposed.