THE PHOTOMULTIPLIER TUBE TESTING FACILITY FOR THE SUDBURY-NEUTRINO-OBSERVATORY

Citation
Cj. Jillings et al., THE PHOTOMULTIPLIER TUBE TESTING FACILITY FOR THE SUDBURY-NEUTRINO-OBSERVATORY, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 373(3), 1996, pp. 421-429
Citations number
12
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
373
Issue
3
Year of publication
1996
Pages
421 - 429
Database
ISI
SICI code
0168-9002(1996)373:3<421:TPTTFF>2.0.ZU;2-T
Abstract
A facility to test the photomultiplier tubes (PMTs) for the Sudbury Ne utrino Observatory (SNO) was constructed at Queen's University. The fa cility measured the noise rate, relative efficiency, and charge and ti me spectra of single photoelectron anode signals for 44 PMTs daily. Th e system is described, and detailed results of the testing for two par ticular PMTs are presented as well as statistics for 9829 Hamamatsu R1 408 PMTs accepted for use in SNO.