Purpose: Placement of a sterile single-use cover over the tonometer ti
p may be a convenient and safe alternative method to repeated chemical
disinfection. This study was undertaken to evaluate the accuracy and
performance of Goldmann applanation tonometry using a sterile disposab
le silicone tonometer shield over the biprism tip. Methods: The same i
nvestigator measured intraocular pressure in 120 eyes with and without
the shield. The eyes were tested randomly first with either the uncov
ered or the covered tonometer to control for the possible effects of r
epeated tonometry influencing measurement differences. Readings were r
ecorded independently in a masked fashion. Results: The intraocular pr
essure measurements evaluated with the silicone shield caused an avera
ge overestimation of the true intraocular pressure of 1.9 +/- 1.9 mmHg
(P = 0.0001). Covered and uncovered readings correlated well (r = 0.9
1, P = 0.0001). The sensitivity and specificity of the covered tonomet
ry in detecting intraocular pressures of 21 mmHg or higher (as measure
d without the shield) were 96.3% and 68.8%, respectively. Minor distor
tions of the normal-appearing fluorescein bands were seen in 5% of the
tonometries, and the shield had to be replaced because of bubbles or
wrinkles on the applanating surface in 3.3% of the measurements, which
means that care must be taken when fitting the shield. Conclusions: A
lthough covered and uncovered tonometry readings correlated well, an o
verall trend for a slightly increased measured intraocular pressure as
a result of using the silicone shield over the Goldmann tonometer tip
was found. The outstanding sensitivity of this procedure in detecting
increased intraocular pressure suggests that it can be useful in scre
ening for ocular hypertension and glaucoma.