Amhp. Vandenbesselaar et Rhm. Peters, MULTICENTER EVALUATION OF THE THROMBOTEST INTERNATIONAL SENSITIVITY INDEX USED WITH A STEEL BALL COAGULOMETER, Journal of Clinical Pathology, 49(5), 1996, pp. 414-417
Aim-To compare the International Sensitivity Index (ISI) of the Thromb
otest reagent used with a steel ball coagulometer (KC) to the ISI of t
he same reagent used with the manual (tilt tube) technique. Methods-Th
e study was carried out by eight laboratories using their own KC instr
ument and method of testing. All laboratories used the same batch of T
hrombotest to determine the clotting times of fresh blood samples from
20 local healthy volunteers and 60 patients on long term oral anticoa
gulant therapy. KC clotting times were plotted against manual clotting
times on double logarithmic scales. Orthogonal regression Lines were
calculated to assess the ISI. Results-In two laboratories the ISI of t
he KC method was lower than that of the manual method; these differenc
es, however, were 2% or less. In the other laboratories no clinically
important differences were observed between ISI values obtained. Howev
er, the clotting times determined with the KC methods were shorter tha
n the manual values. Conclusions-The ISI of Thrombotest determined wit
h the KC methods was very similar to the manual value. Therefore, use
of the ISI value supplied by the manufacturer without adjustment is ju
stified. The mean normal prothrombin time, however, must be determined
locally.