Jw. Mclaren et al., ONLINE METHOD FOR THE ANALYSIS OF SEA-WATER FOR TRACE-ELEMENTS BY INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY, Journal of analytical atomic spectrometry, 8(2), 1993, pp. 279-286
A commercially available chelation concentration system for the on-lin
e preconcentration of trace metals from saline solutions was modified
to permit the simultaneous determination of Fe, Mn, Co, Ni, Cu, Zn, Cd
and Pb in 5 ml sea-water samples in less than 10 min by inductively c
oupled plasma mass spectrometry. Modifications were necessary in order
to obtain acceptably low blanks for several elements. The system was
operated both with columns provided by the manufacturer and with colum
ns containing silica-immobilized 8-hydroxyquinoline. Detection limits
in sea-water ranged from 1.6 ng l-1 (for Pb) to 55 ng l-1 (for Ni) for
the former columns, and from 0.3 ng l-1 (for Cd) to 47 ng l-1 (for Fe
) for the latter. Accurate analyses of coastal and open-ocean sea-wate
r certified reference materials were achieved with at least one of the
column types for all elements except Fe and Co.