BODY FIXED ANKLE DYNAMOMETER FOR ISOMETRIC TESTS IN-SPACE FLIGHTS

Citation
D. Rafolt et E. Gallasch, BODY FIXED ANKLE DYNAMOMETER FOR ISOMETRIC TESTS IN-SPACE FLIGHTS, Biomedizinische Technik, 41(4), 1996, pp. 91-97
Citations number
9
Categorie Soggetti
Engineering, Biomedical","Medical Informatics
Journal title
ISSN journal
00135585
Volume
41
Issue
4
Year of publication
1996
Pages
91 - 97
Database
ISI
SICI code
0013-5585(1996)41:4<91:BFADFI>2.0.ZU;2-7
Abstract
To study neuromuscular variables in isometric contraction, a body-fixe d ankle dynamometer with integrated electromyographic (EMG) system and software-controlled load tests was developed. It consists of a footpl ate acting against a load cell in plantar flexion. Counteraction is su pported by a perpendicular telescopic stand which runs from the heel t o the thigh with the knee flexed at an angle of 90 degrees. Additional to contraction force, force tremor and surface EMG from the triceps s urae and tibialis anterior muscles are recorded. The software consists of three load tests to assess variables of muscle fatigue, motor perf ormance and motor memory. The dynamometer was optimised in relation to functionality and selection of construction materials for the use on the Russian MIR space station.