DETERMINATION OF DILTIAZEM HYDROCHLORIDE IN HUMAN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY

Citation
Rs. Chaudhary et al., DETERMINATION OF DILTIAZEM HYDROCHLORIDE IN HUMAN SERUM BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY, Journal of chromatography. Biomedical applications, 614(2), 1993, pp. 261-266
Citations number
25
Categorie Soggetti
Chemistry Analytical
ISSN journal
03784347
Volume
614
Issue
2
Year of publication
1993
Pages
261 - 266
Database
ISI
SICI code
0378-4347(1993)614:2<261:DODHIH>2.0.ZU;2-Y
Abstract
A simple and sensitive reversed-phase high-performance liquid chromato graphic method for the determination of diltiazem in human serum has b een developed. The method involves a one-step deproteinization of seru m for sample clean-up using acetonitrile. A LiChrosorb RP-8 column (30 cm x 4.1 mm I.D.) was eluted isocratically with acetonitrile-0.01 M d ibasic sodium phosphate (40:60, v/v) containing 0.01% triethanolamine. Diltiazem was monitored at 237 nm and 0.1 a.u.fs. The completion time for assay was less than 15 min, and the lower limit of quantitation w as 10 ng/ml for a 100-mul injection volume. Using this method, the pha rmacokinetic parameters were calculated from a serum concentration ver sus time profile of diltiazem.