PROCEDURE FOR MEASURING THE SENSITIVITY OF PRECISION DETECTORS OF VISIBLE AND INFRARED SYNCHROTRON-RADIATION

Citation
Aa. Maltsev et Ma. Maltsev, PROCEDURE FOR MEASURING THE SENSITIVITY OF PRECISION DETECTORS OF VISIBLE AND INFRARED SYNCHROTRON-RADIATION, Measurement techniques, 38(10), 1995, pp. 1194-1198
Citations number
12
Journal title
ISSN journal
05431972
Volume
38
Issue
10
Year of publication
1995
Pages
1194 - 1198
Database
ISI
SICI code
0543-1972(1995)38:10<1194:PFMTSO>2.0.ZU;2-N
Abstract
A procedure is described for measuring the total sensitivity of photod etectors used in precision measurements of the number of low-energy (E < 20 MeV) relativistic electrons from the intensity of their synchrot ron radiation in the visible and infrared regions of the spectrum. The calibration constants of two types of uncooled photodetectors (a lead selenide photoresistor and a silicon photodiode) are measured and the measurement errors are analyzed.