DESIGN CONSIDERATIONS FOR CAPILLARY-FLOW INJECTION SYSTEMS

Citation
Dm. Spence et al., DESIGN CONSIDERATIONS FOR CAPILLARY-FLOW INJECTION SYSTEMS, Instrumentation science & technology, 24(2), 1996, pp. 103-113
Citations number
9
Categorie Soggetti
Instument & Instrumentation","Chemistry Analytical
ISSN journal
10739149
Volume
24
Issue
2
Year of publication
1996
Pages
103 - 113
Database
ISI
SICI code
1073-9149(1996)24:2<103:DCFCIS>2.0.ZU;2-M
Abstract
Design considerations for performing capillary now injection (CFI) wit h conventional peristaltic pumping are presented. A time based injecti on technique has been developed so that sample volumes in the sub-nL t o mt range can be used without internal or external injection loops. I n addition, two now cells have been designed and constructed in order to measure absorbances with such small sample volumes. Commercially av ailable connectors for fused silica capillaries have also been used in a CFI system enabling the operation of multiple reagent systems.