MEASUREMENT OF DEVIATION OF THE MICROSCOPE STAGE FROM SQUARENESS TO THE OPTICAL-AXIS

Authors
Citation
O. Prakash et Rs. Ram, MEASUREMENT OF DEVIATION OF THE MICROSCOPE STAGE FROM SQUARENESS TO THE OPTICAL-AXIS, Journal of optics, 27(1), 1996, pp. 31-33
Citations number
3
Categorie Soggetti
Optics
Journal title
ISSN journal
0150536X
Volume
27
Issue
1
Year of publication
1996
Pages
31 - 33
Database
ISI
SICI code
0150-536X(1996)27:1<31:MODOTM>2.0.ZU;2-H
Abstract
The measurement of deviation of microscope stage from squareness to th e optical axis is an important parameter in optical testing of microsc opes. This test ensures that when the deviation is in the permitted li mits, the optical axis of the microscope is normal to the stage and he nce the movement of the microscope tube is in normal direction to the stage. In the present article a simple method to measure the deviation of the microscope stage from squareness to its optical axis has been described using a parallel plate and autocollimators. Test results on a few microscopes have been presented.