Mn. Yongbi et al., A MODIFIED SUB-SECOND FAST-STEAM SEQUENCE INCORPORATING BIPOLAR GRADIENTS FOR IN-VIVO DIFFUSION IMAGING, Magnetic resonance in medicine, 35(6), 1996, pp. 911-916
A modified high-speed stimulated-echo acquisition mode (STEAM) diffusi
on sequence (90 degrees-TE/2-90 degrees-TM-[alpha-TE/2-STE](n)) incorp
orating bipolar diffusion gradients that are less sensitive to macrosc
opic motion-induced artifacts is presented, Diffusion encoding was per
formed only during the first echo interval (TE1) with bipolar gradient
s that were implemented on all three mutually orthogonal axes. Calibra
tion measurements on phantoms filled with water, isopropanol, and dime
thyl sulfoxide yielded apparent diffusion coefficients (ADC) consisten
t with published values, Non-EGG-triggered in vivo images acquired on
rat brain with relatively high b values (similar to 450 s/mm(2)) indic
ated minimal motion artifacts. Evaluated ADCs averaged over the cortex
, left mid-brain, right mid-brain, regions yielded (0.91 +/- 0.02), (1
.06 +/- 0.02), (1.01 +/- 0.03) x 10(-3) mm(2)s(-1), respectively.