ZEOLITE-CONTAINING OXIDASE-BASED CARBON-PASTE BIOSENSORS

Citation
J. Wang et A. Walcarius, ZEOLITE-CONTAINING OXIDASE-BASED CARBON-PASTE BIOSENSORS, Journal of electroanalytical chemistry [1992], 404(2), 1996, pp. 237-242
Citations number
9
Categorie Soggetti
Electrochemistry,"Chemistry Analytical
Journal title
Journal of electroanalytical chemistry [1992]
ISSN journal
15726657 → ACNP
Volume
404
Issue
2
Year of publication
1996
Pages
237 - 242
Database
ISI
SICI code
Abstract
The incorporation of zeolite particles within glucose oxidase containi ng carbon paste leads to improved amperometric biosensing of glucose. Such zeolite dispersion offers enhanced sensitivity, extended linear r ange and higher stability. Such improvements are attributed to the for mation of a porous paste network that exposes the enzyme contained in the interior of the paste to the substrate solution and imparts higher mass transport resistance. Several electrode configurations shed usef ul insights into the diffusion processes within the compact zeolite-ca rbon paste-enzyme microstructure and the thickness of the active enzym e layer. The mass-transport induced increases in K-m are attributed to diffusional limitations within the electrode interior.