IMPROVED ALGORITHM FOR LINEARIZATION OF CALIBRATION CURVES IN ZEEMAN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY

Citation
Bv. Lvov et al., IMPROVED ALGORITHM FOR LINEARIZATION OF CALIBRATION CURVES IN ZEEMAN GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY, Spectrochimica acta, Part B: Atomic spectroscopy, 51(6), 1996, pp. 609-618
Citations number
11
Categorie Soggetti
Spectroscopy
ISSN journal
05848547
Volume
51
Issue
6
Year of publication
1996
Pages
609 - 618
Database
ISI
SICI code
0584-8547(1996)51:6<609:IAFLOC>2.0.ZU;2-0
Abstract
A two-parameter model for simulation of concentration curves in Zeeman graphite furnace atomic absorption spectrometry is proposed. The algo rithm based on this model can be used for linearization of calibration curves up to the rollover point. The merits of the algorithm are supp orted by experimental data obtained for 20 elements under different me asurement conditions (light source current and slit width), including the cases where the curvature of the initial calibration curves origin ates partially from mass-dependent chemical effects and/or non-uniform atom distribution over the furnace cross section. The linearization e rror does not exceed the random scatter for replicates.