INFLUENCE OF THE SOLID-PHASE EXTRACTION PROCESS ON CALIBRATION AND PERFORMANCE PARAMETERS FOR THE DETERMINATION OF PESTICIDE-RESIDUES IN WATER BY GAS-CHROMATOGRAPHY
C. Delacolina et al., INFLUENCE OF THE SOLID-PHASE EXTRACTION PROCESS ON CALIBRATION AND PERFORMANCE PARAMETERS FOR THE DETERMINATION OF PESTICIDE-RESIDUES IN WATER BY GAS-CHROMATOGRAPHY, Journal of chromatography, 733(1-2), 1996, pp. 275-281
Citations number
21
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
Twenty-six organophosphorus, organochlorine and other electron-capture
detection-sensitive pesticides were extracted from water in a single
step using a C-18 solid-phase extraction cartridge, eluted with ethyl
acetate and isooctane and determined by gas chromatography with electr
on-capture and flame photometric detection. The calibration equation f
or the extraction method was calculated for a twentyfold concentration
range, including the EC limit of 0.1 mu g/l. The linearity, precision
, sensitivity and detection limit of the method were studied, applying
the statistical model of linear regression. A lack of linearity was o
bserved for fenthion, deltamethrin and trifluralin, but the proposed m
ethod was suitable for other pesticides studied. The limits of detecti
on range from 20 to 120 ng II applying the calibration graph and from
1 to 40 ng/l based on a signal-to-noise ratio of 3:1.