Objectives. The aim of tills study was the assessment of the duality o
f side-firing fibers that are being used for laser prostatectomy, eith
er by a laser light transmission measurement or by visual inspection.
Methods. A power meter (Aquarius) was developed to measure the actual
power transmitted through a side-firing fiber and delivered to the pro
static tissue. The power measurements were performed under clinical co
nditions, that is, under water and at relatively high input power. Fur
thermore, a protocol was developed for visual inspection of the fibers
. Eight types of side-firing fibers were measured before use, Before a
nd after a procedure, three fiber types were measured: ProLase II (28
samples), UltraLine (23 samples), and UroLase (44 samples). Ail these
fibers were used in standard treatment protocols. Results. At 60 W the
transmission ai new fibers (not used) ranged between 49% and 83% when
compared to a bare fiber. After use, a large variation was found in t
ransmitted power between different samples of one device. A correlatio
n with total transmitted power was not present. At higher power input,
vapor bubbles are generated at the tip of the fibers. Depending on th
e fiber design, these bubbles have a major impact on the transmission.
Only for the UroLase fiber was there a significant correlation betwee
n visual inspection and the transmission of used samples at 10, 20, an
d 40 W. Conclusions. The transmission strongly varies between fibers a
nd between different samples of one fiber during clinical use. Moreove
r, the transmission does not correlate with visual inspection, A power
measurement during a clinical treatment will contribute to a more con
trolled procedure and to a better comparison of clinical laser prostat
ectomy studies.