A RAPID ANION-EXCHANGE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD FOR THE MEASUREMENT OF HBA(2) IN WHOLE-BLOOD

Citation
Bg. Keevil et al., A RAPID ANION-EXCHANGE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY METHOD FOR THE MEASUREMENT OF HBA(2) IN WHOLE-BLOOD, Annals of clinical biochemistry, 33, 1996, pp. 253-256
Citations number
7
Categorie Soggetti
Biology,"Chemistry Medicinal
ISSN journal
00045632
Volume
33
Year of publication
1996
Part
3
Pages
253 - 256
Database
ISI
SICI code
0004-5632(1996)33:<253:ARAHLM>2.0.ZU;2-9
Abstract
We developed a binary gradient high-performance liquid chromatography (HPLC) method for measuring HbA(2) in whole blood samples using a Phar macia Mono Q column (1 mt) and measurement at 415 nm. The assay requir es a simple lysis and centrifugation step before injection onto the co lumn. We found good agreement of results between the HPLC method and t he Helena column chromatography method. The within batch precision was 2.6% and between batch precision was 4.6%. We found that using 30 mM Tris buffers (pH 7.8) with a sodium chloride gradient resulted in shor t analysis times and good chromatographic separation of HbA(2), HbS an d HbA. We conclude that this is a robust assay for the diagnosis of be ta-thalassaemia.