REPORT ON COOPERATIVE DETERMINATION OF MO LECULAR-WEIGHT AVERAGES OF POLYMERS BY SIZE-EXCLUSION CHROMATOGRAPHY .3. ON THE POLYMER STANDARDSFOR CALIBRATION
S. Mori et al., REPORT ON COOPERATIVE DETERMINATION OF MO LECULAR-WEIGHT AVERAGES OF POLYMERS BY SIZE-EXCLUSION CHROMATOGRAPHY .3. ON THE POLYMER STANDARDSFOR CALIBRATION, Bunseki Kagaku, 45(5), 1996, pp. 447-453
The weight average molecular weight (M(W)) of polystyrene samples dete
rmined in the first round robin test (RR-1) was compared with those in
the second round robin test (RR-2) to analyse the difference in polys
tyrene standards for calibration. The data were divided into two group
s: data determined at laboratories that used the standards obtained by
the identical vendor in both RRs (case 1) and those using the standar
ds of different vendors for RR-1 and RR-2 (case 2). The RR-1: RR-2 rat
ios for M(W) in case 1 were 1.03 similar to 1.04 in average, but the m
aximum was 1.17 and the minimum 0.95. By careful adjustment of calibra
tion conditions, the RR-1: RR-2 ratios for M(W) were reduced to 1.01 s
imilar to 1.03. In case 2, as long as a calibration was constructed by
passing through most data points smoothly, there was no distinct diff
erence in calculated M(W) among standards from different vendors. The
relative standard deviation of these data, obtained carefully as menti
oned above, were 3.9%.