ION-SELECTIVE ELECTRODE FOR SALICYLATE ASSAY IN BLOOD-SERUM

Authors
Citation
T. Katsu et Y. Mori, ION-SELECTIVE ELECTRODE FOR SALICYLATE ASSAY IN BLOOD-SERUM, Talanta, 43(5), 1996, pp. 755-759
Citations number
19
Journal title
Talanta
ISSN journal
00399140 → ACNP
Volume
43
Issue
5
Year of publication
1996
Pages
755 - 759
Database
ISI
SICI code
0039-9140(1996)43:5<755:IEFSAI>2.0.ZU;2-3
Abstract
A salicylate-selective membrane electrode made with heptyl-4-trifluoro acetylbenzoate as a neutral carrier was successfully applied for the d etermination of salicylate in blood serum. This procedure is advantage ous because the free concentration of the drug in serum can be determi ned without sample preparation. The free salicylate concentrations det ermined-by the ion-selective electrode compared to those obtained by c onventional colorimetry gave a linear correlation coefficient 0.997 in the salicylate concentration range 0.1-2.5 mM.