St. Woolson et Cg. Comstock, POROUS PAD SEPARATION AND LOOSENING OF HARRIS-GALANTE FEMORAL HIP COMPONENTS, The Journal of arthroplasty, 11(4), 1996, pp. 474-477
Failure of the bond between the metal beads and the solid substrate of
porous-coated chrome-cobalt femoral components, or so-called ''bead s
hedding,'' has been reported previously and may indicate motion of the
component within the femur or mechanical loosening of the prosthesis.
Failure of the bond between a sintered titanium fibermesh pad and a t
itanium alloy femoral stem has not been documented in the literature t
o date. The separation of fibermesh Dads from three Harris-Galante fem
oral prostheses (Zimmer, Warsaw, IN) is reported.