THE NONPROPORTIONAL RESPONSE OF SINGLE-CRYSTALLINE OXIDE SCINTILLATORS

Citation
Ep. Sysoeva et al., THE NONPROPORTIONAL RESPONSE OF SINGLE-CRYSTALLINE OXIDE SCINTILLATORS, IEEE transactions on nuclear science, 43(3), 1996, pp. 1282-1283
Citations number
5
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
43
Issue
3
Year of publication
1996
Part
2
Pages
1282 - 1283
Database
ISI
SICI code
0018-9499(1996)43:3<1282:TNROSO>2.0.ZU;2-7
Abstract
The energy dependence of the response of single crystalline oxide scin tillators (EGO, CWO and GSO(Ce)) was studied over the wide energy rang e from 6 to 1332 keV. The non-proportionality was defined as the diffe rence between the specific response value at the given energy and the one at 661.6 keV. It is found that a considerable (up to 35%) non-prop ortionality of the response exists for the all investigated scintillat ors in the low energy region from 6 to 100 keV, while as in the energy range from 100 to 1332 keV the scintillator's response is proportiona l to the gamma-rays energy with the accuracy equal to +/- 1%.