A SIMPLE METHOD TO IMPROVE THE SPATIAL UNIFORMITY OF VENETIAN-BLIND PHOTOMULTIPLIER TUBES

Citation
Jmf. Dossantos et al., A SIMPLE METHOD TO IMPROVE THE SPATIAL UNIFORMITY OF VENETIAN-BLIND PHOTOMULTIPLIER TUBES, IEEE transactions on nuclear science, 43(3), 1996, pp. 1335-1340
Citations number
13
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
43
Issue
3
Year of publication
1996
Part
2
Pages
1335 - 1340
Database
ISI
SICI code
0018-9499(1996)43:3<1335:ASMTIT>2.0.ZU;2-7
Abstract
An improvement in the uniformity of venetian-blind photomultiplier tub es has been achieved by reducing the voltage difference between the fi rst and second dynodes. The method has been applied to a gas proportio nal scintillation counter (GPSC) instrumented with a venetian-blind ph otomultiplier (PMT). When exposed to a 20-mm collimated 5.9-keV x-ray beam, an overall improvement in energy resolution for the GPSC/PMT com bination from 20% to 11.5% was achieved. An alternative method that re duces the photocathode-to-first-dynode voltage was less effective and resulted in a severe degradation of detector energy resolution.