Jmf. Dossantos et al., A SIMPLE METHOD TO IMPROVE THE SPATIAL UNIFORMITY OF VENETIAN-BLIND PHOTOMULTIPLIER TUBES, IEEE transactions on nuclear science, 43(3), 1996, pp. 1335-1340
An improvement in the uniformity of venetian-blind photomultiplier tub
es has been achieved by reducing the voltage difference between the fi
rst and second dynodes. The method has been applied to a gas proportio
nal scintillation counter (GPSC) instrumented with a venetian-blind ph
otomultiplier (PMT). When exposed to a 20-mm collimated 5.9-keV x-ray
beam, an overall improvement in energy resolution for the GPSC/PMT com
bination from 20% to 11.5% was achieved. An alternative method that re
duces the photocathode-to-first-dynode voltage was less effective and
resulted in a severe degradation of detector energy resolution.