Thvt. Dias et al., THE RESPONSE OF XENON X-RAY-DETECTORS TO FULL-ENERGY ABSORPTION AND FLUORESCENCE-ESCAPE EVENTS - MEASUREMENT AND MODELING, IEEE transactions on nuclear science, 43(3), 1996, pp. 1432-1441
We have observed that the pulse-height distributions of x-ray events t
hat deposit essentially the same amount of energy in a xenon gas detec
tor will vary according to the deexcitation pathways available to the
photoionized atoms. A xenon gas proportional scintillation counter was
used to compare the pulse-height distributions of events in which flu
orescence x-rays from xenon escape detection. Experimental results are
presented and the behaviour for different types of events is clarifie
d using Monte Carlo simulation. The results indicate that particular c
are must be taken when using escape peaks for energy calibration of xe
non gas detectors.