BASIC PERFORMANCE OF UNITIZED COMPTON-SCATTERING TYPE POLARIMETER

Citation
H. Tomita et al., BASIC PERFORMANCE OF UNITIZED COMPTON-SCATTERING TYPE POLARIMETER, IEEE transactions on nuclear science, 43(3), 1996, pp. 1527-1532
Citations number
15
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
43
Issue
3
Year of publication
1996
Part
2
Pages
1527 - 1532
Database
ISI
SICI code
0018-9499(1996)43:3<1527:BPOUCT>2.0.ZU;2-F
Abstract
We have been developing a Compton scattering type polarimeter which co nsists of nine unit counters to detect the polarization of hard X-rays from stellar objects such as the Crab Nebula. We constructed one unit counter which consists of one plastic scintillator(7x7x5cm(3)) and fo ur phoswich scintillators(6.7x1.1x32.2cm(3)) made of NaI(Tl)+CsI(Na). The characteristics of the unit counter were measured in detail to det ermine the detection efficiency (20.3%) and the modulation factor (27. 5%) at 53.4keV. The background spectrum of the unit counter in the lab oratory is also measured. Based on the results, the background from th e outside of the active collimator is reduced by 50.5% when used with the PSD. On the basis of these results, the performance of the planned polarimeter was estimated by computer simulations. The results showed that the polarimeter can detect the polarization of hard X-rays from the Crab Nebula at a MDP (Minimum Detectable Polarization) of 14.0% wi th a 3-hour balloon observation.