SMALL CRYOSTEM FOR OPERATION OF VISIBLE-LIGHT PHOTON COUNTERS (VLPC)

Citation
B. Baumbaugh et al., SMALL CRYOSTEM FOR OPERATION OF VISIBLE-LIGHT PHOTON COUNTERS (VLPC), IEEE transactions on nuclear science, 43(3), 1996, pp. 1741-1745
Citations number
7
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
43
Issue
3
Year of publication
1996
Part
2
Pages
1741 - 1745
Database
ISI
SICI code
0018-9499(1996)43:3<1741:SCFOOV>2.0.ZU;2-7
Abstract
We have designed, constructed, and operated a small cryostem which sup ports the operation of 32 channels of Visible Light Photon Counters (H ISTE-IV VLPCs). The VLPCs are situated within a small enclosure which can be lowered into a 30-100 liter dewar. The enthalpy of the boil-off helium keeps the VLPCs cold, and allows the system to be operated wit h stability for many days within the desired temperature range of 6.0K -7.5K. The cryostem is instrumented with clear fiber waveguides which transport the light from an optical connector situated at the top of t he cryostem and outside of the dewar to the photosensors at cryogenic temperatures within the dewar. Electrical signals from the VLPCs are a mplified at room temperature using QPA02 preamplifiers. Details of des ign and performance are reviewed.