C. Moisan et al., PERFORMANCE STUDIES OF A DEPTH ENCODING MULTICRYSTAL DETECTOR FOR PET, IEEE transactions on nuclear science, 43(3), 1996, pp. 1926-1931
We report the results of design studies for a multicrystal detector wi
th the capacity to encode the Depth-Of-Interaction (DOI). These studie
s are based on a simulation that treats the interactions of gamma-rays
and tracks scintillation photons in state-of-the-art position encodin
g PET detectors. The simulation was used to study the impact of a simp
le modification to the standard surface treatment of the EXACT HR PLUS
block that induces a significant sensitivity of the photopeak pulse h
eight upon the DOI of 511 keV gamma-rays. The depth resolution achieve
d with the altered block as well as the impact of the modification on
its energy and position resolutions are presented. The results show a
depth resolution of 6 to 7 mm for the inner crystals of the modified b
lock. The center crystals show the largest increase in event position
miscoding with 16% less encoding accuracy in the modified block compar
ed to the original one.