PERFORMANCE STUDIES OF A DEPTH ENCODING MULTICRYSTAL DETECTOR FOR PET

Citation
C. Moisan et al., PERFORMANCE STUDIES OF A DEPTH ENCODING MULTICRYSTAL DETECTOR FOR PET, IEEE transactions on nuclear science, 43(3), 1996, pp. 1926-1931
Citations number
9
Categorie Soggetti
Nuclear Sciences & Tecnology","Engineering, Eletrical & Electronic
ISSN journal
00189499
Volume
43
Issue
3
Year of publication
1996
Part
2
Pages
1926 - 1931
Database
ISI
SICI code
0018-9499(1996)43:3<1926:PSOADE>2.0.ZU;2-J
Abstract
We report the results of design studies for a multicrystal detector wi th the capacity to encode the Depth-Of-Interaction (DOI). These studie s are based on a simulation that treats the interactions of gamma-rays and tracks scintillation photons in state-of-the-art position encodin g PET detectors. The simulation was used to study the impact of a simp le modification to the standard surface treatment of the EXACT HR PLUS block that induces a significant sensitivity of the photopeak pulse h eight upon the DOI of 511 keV gamma-rays. The depth resolution achieve d with the altered block as well as the impact of the modification on its energy and position resolutions are presented. The results show a depth resolution of 6 to 7 mm for the inner crystals of the modified b lock. The center crystals show the largest increase in event position miscoding with 16% less encoding accuracy in the modified block compar ed to the original one.