MEASUREMENT OF WEAK INTRACAVITY ANISOTROP IES USING HIGH-FINESSE FABRY-PEROT CAVITIES

Citation
J. Poirson et al., MEASUREMENT OF WEAK INTRACAVITY ANISOTROP IES USING HIGH-FINESSE FABRY-PEROT CAVITIES, Annales de physique, 20(5-6), 1995, pp. 607-608
Citations number
6
Categorie Soggetti
Physics
Journal title
ISSN journal
00034169
Volume
20
Issue
5-6
Year of publication
1995
Pages
607 - 608
Database
ISI
SICI code
0003-4169(1995)20:5-6<607:MOWIAI>2.0.ZU;2-Q
Abstract
The dependence of the eigenstates of a high-finesse Fabry-Perot is sho wn to provide a very sensitive method to characterize residual anisotr opies of supemirrors at normal incidence and to measure weak intracavi ty anisotropies. It leads to a sensitivity of. 10(-10) rad for phase r etardance detection. A precise method for large finesse measurements i s also demonstrated.