Vs. Burakov et al., DETERMINATION OF CESIUM, MANGANESE AND TERBIUM BY INTRACAVITY LASER SPECTROSCOPY, Fresenius' journal of analytical chemistry, 355(3-4), 1996, pp. 317-318
The statistical characteristics of the intracavity laser spectrometer
are investigated by analysis of cesium in aqueous solutions. The error
distribution of measurements is normal. The relative standard deviati
on of the analysis in the middle of the dynamic range is equal to 9.7%
. The intracavity spectral analysis was carried out at low contents of
cesium: manganese and terbium in aqueous solutions. The limits of det
ection are: 5 mu g l(-1) (Cs), 20 mu g l(-1) (Mn) and 1000 mu g l(-1)
(Tb). Owing to the high spectral resolution of the spectrometer used t
he hyperfine structure of the thallium absorption line (535.05 nm) was
recorded.