T. Saito et H. Onuki, DETECTOR CALIBRATION IN THE WAVELENGTH REGION 10-NM TO 100-NM BASED ON A WINDOWLESS RARE-GAS IONIZATION-CHAMBER, Metrologia, 32(6), 1996, pp. 525-529
A detector calibration facility based on a rare gas ionization chamber
has been built at the Electrotechnical Laboratory, Japan. We have rea
lized a windowless system by applying strong differential pumping, whi
ch features no spectral limitation and easy gas handling. In the auto-
ionization region near the xenon absorption edge, photon flux was foun
d to be underestimated and this limited the maximum available waveleng
th to 92 nm. The uncertainty (1 sigma) for a Si photodiode calibration
was estimated to be within +/-9% in the wavelength range 10 nm to 92
nm.