RADIOMETRIC REALIZATION OF THE CANDELA WITH A TRAP DETECTOR

Citation
E. Ikonen et al., RADIOMETRIC REALIZATION OF THE CANDELA WITH A TRAP DETECTOR, Metrologia, 32(6), 1996, pp. 689-692
Citations number
11
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
Journal title
ISSN journal
00261394
Volume
32
Issue
6
Year of publication
1996
Pages
689 - 692
Database
ISI
SICI code
0026-1394(1996)32:6<689:RROTCW>2.0.ZU;2-E
Abstract
We have developed measurement systems for radiometric realization of t he units of luminous intensity and illuminance. The advantages of the trap-detector-based systems include a low level of detector-filter int erreflections and good spatial homogeneity. Evaluation of known uncert ainty components gives a total uncertainty of 0,6% (2 sigma).