CALIBRATION OF THE BROAD-BAND RADIOMETERS OF THE FINNISH SOLAR ULTRAVIOLET MONITORING NETWORK

Citation
K. Leszczynski et al., CALIBRATION OF THE BROAD-BAND RADIOMETERS OF THE FINNISH SOLAR ULTRAVIOLET MONITORING NETWORK, Metrologia, 32(6), 1996, pp. 701-704
Citations number
11
Categorie Soggetti
Physics, Applied","Instument & Instrumentation
Journal title
ISSN journal
00261394
Volume
32
Issue
6
Year of publication
1996
Pages
701 - 704
Database
ISI
SICI code
0026-1394(1996)32:6<701:COTBRO>2.0.ZU;2-G
Abstract
The present solar ultraviolet (UV) monitoring network of the Finnish M eteorological Institute consists of seven sites equipped with erythema lly-weighted broadband radiometers. Two of the sites are also equipped with a Brewer spectroradiometer. The Finnish Centre for Radiation and Nuclear Safety is responsible for testing and calibration of the broa dband radiometers. An Optronic 742 spectroradiometer is used as a refe rence instrument, and the broadband radiometers are calibrated against the Optronic 742 in solar radiation. The Optronic 742 is calibrated a gainst 1000 W FEL lamps traceable to the NIST. The errors caused by no nideal radiometric characteristics of the Optronic 742 are numerically corrected. The uncertainty (2 sigma) of solar UV measurements with th e Optronic 742 is estimated to be +/- 8 % at 310 nm and the uncertaint y of the spectroradiometric solar calibration of the Solar Light Model 501 radiometers is estimated to be +/- 11%.