A FIDELITY CHECK OF A PRACTICAL CONE-BEAM TOMOGRAPHY ALGORITHM

Citation
P. Munshi et al., A FIDELITY CHECK OF A PRACTICAL CONE-BEAM TOMOGRAPHY ALGORITHM, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 111(3-4), 1996, pp. 337-340
Citations number
21
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
111
Issue
3-4
Year of publication
1996
Pages
337 - 340
Database
ISI
SICI code
0168-583X(1996)111:3-4<337:AFCOAP>2.0.ZU;2-J
Abstract
Cone-beam tomography is currently being developed for commercial appli cations wherever two-dimensional X-ray tomography is being used. One p opular method for reconstructing a three dimensional image of an objec t (from cone-beam data), is the Feldkamp-Davis-Kress (FDK) algorithm, and it is an extension of the well known convolution and backprojectio n (CBP) method used in the usual two dimensional cases. This study inv estigates the fidelity of approximations involved in the FDK algorithm . Test data is obtained by the cone-beam X-ray scanner developed at th e Institute of Cancer Research (Surrey, England). The error formula fo r the CBP filters plays a key role in this work.