EXPERIMENTAL X-RAY PEAK-SHAPE DETERMINATION FOR A SI(LI) DETECTOR

Citation
Lc. Alves et al., EXPERIMENTAL X-RAY PEAK-SHAPE DETERMINATION FOR A SI(LI) DETECTOR, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 109, 1996, pp. 129-133
Citations number
14
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
109
Year of publication
1996
Pages
129 - 133
Database
ISI
SICI code
0168-583X(1996)109:<129:EXPDFA>2.0.ZU;2-C
Abstract
The response function of a Si(Li) detector to monochromatic radiation was determined in the energy region between 1.7 and 10 keV A wavelengt h dispersive spectrometer with a plane Bragg crystal was used to selec t the X-ray lines to be analysed. The chosen targets were irradiated w ith 1.8 MeV protons. Some modifications were introduced to the known H YPERMET function in order to describe the peak shapes observed. An ene rgy dependent function was then determined involving the parameters us ed in the detector response function. The energy dispersive analysis o f the Mn K radiation resulting from the decay of a Fe-55 radioactive s ource, revealed additional radiative contributions to the peak shape o bserved, attributed to the Auger radiative effect and to the Mn K-beta 5 line.