Ajm. Plompen et al., LOW CURRENT CHARGE NORMALIZATION WITH SCINTILLATORS, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 109, 1996, pp. 161-166
The use of ZnS(Ag) and YAG(Ce) scintillators for charge normalization
at low currents (>15 pA) was demonstrated with current gains of 2 x 10
(3) using a photomultiplier tube (PMT) and up to 10 with a photodiode
(PD). An accuracy of a few percent, sufficient for ion beam applicatio
ns, was obtained. The potential of the method and its use for micropro
be applications is discussed.