FEASIBILITY STUDY OF A PORTABLE PIXE SYSTEM USING A PO-210 ALPHA-SOURCE

Citation
G. Pappalardo et al., FEASIBILITY STUDY OF A PORTABLE PIXE SYSTEM USING A PO-210 ALPHA-SOURCE, Nuclear instruments & methods in physics research. Section B, Beam interactions with materials and atoms, 109, 1996, pp. 214-217
Citations number
5
Categorie Soggetti
Physics, Nuclear","Nuclear Sciences & Tecnology","Instument & Instrumentation
ISSN journal
0168583X
Volume
109
Year of publication
1996
Pages
214 - 217
Database
ISI
SICI code
0168-583X(1996)109:<214:FSOAPP>2.0.ZU;2-P
Abstract
An alpha source which consists of 1 mCi of Po-210 electro-deposited on 1 mu m thin Ag foil attached to a ceramic backing has been prepared. The overall geometry exhibits a truncated-conical shape. The source ha s been coupled to a standard 10 mm diameter Si(Li) X-rays detector con nected to a portable electronic system (ORTEC-NOMAD), The analysed are a of the exposed samples is 16 mm in diameter. Preliminary results sta te that medium-light elements, like Ca, Ti, Fe, Co, Cr, Ni and Cu can be easily analysed at concentrations of some per cent. Lighter element s determination is affected by the enhancing presence of X-rays genera ted in the Ag foil and by some chlorine impurities. PIXE measurements of attic greek pottery sherds are discussed.