MODIFIED ERG-JET CONTACT-LENS ELECTRODES FOR USE IN INFANTS AND TODDLERS

Citation
Se. Brodie et Me. Breton, MODIFIED ERG-JET CONTACT-LENS ELECTRODES FOR USE IN INFANTS AND TODDLERS, Documenta ophthalmologica, 91(2), 1995, pp. 141-146
Citations number
2
Categorie Soggetti
Ophthalmology
Journal title
ISSN journal
00124486
Volume
91
Issue
2
Year of publication
1995
Pages
141 - 146
Database
ISI
SICI code
0012-4486(1995)91:2<141:MECEFU>2.0.ZU;2-Q
Abstract
A simple modification of the standard ERG-jet contact lens recording e lectrode for use in infants and small children was developed. A small Plexiglas cylinder is fixed to the front surface of the electrode, whi ch prevents lid closure and facilitates handling. Sedation or general. anesthesia is rarely needed. Waveforms obtained with the modified ele ctrode appear comparable to those obtained with standard electroretino gram electrodes.