TECHNICAL MODIFICATION OF THE RHEOLYTIC T HROMBECTOMY

Citation
S. Mullerhulsbeck et al., TECHNICAL MODIFICATION OF THE RHEOLYTIC T HROMBECTOMY, RoFo. Fortschritte auf dem Gebiete der Rontgenstrahlen und der neuenbildgebenden Verfahren, 164(6), 1996, pp. 507-510
Citations number
14
Categorie Soggetti
Radiology,Nuclear Medicine & Medical Imaging
ISSN journal
09366652
Volume
164
Issue
6
Year of publication
1996
Pages
507 - 510
Database
ISI
SICI code
0936-6652(1996)164:6<507:TMOTRT>2.0.ZU;2-9
Abstract
A 48-year-old male patient with acute oc elusion (5 cm) of a leg arter y was treated with the rheolytic thrombectomy (RT). As recommended the catheter is applicated via a single coaxial running guide wire, which is removed later. If an additional wire has been placed in the occlud ed vessel segment, the occlusion can be passed repeatedly. A 30 degree s bent catheter tip improves the removal of wall-adherent thrombotic m aterial. The occluded segment was successfully recanalised with the RT C, 270 ml of heparinised saline had been injected. The blood examinati ons showed mild haemolysis.