Failure analysis of a retrieved flexible intramedullary nail was perfo
rmed by standard optical and scanning electron microscopy, Results sho
w that the corrosion fatigue at the stamped label caused the nail frac
ture, The authors strongly recommend that the American Society for Tes
ting of Materials standard F86-91 be modified to exclude mechanical st
amping from the repertoire for labeling implants, Surgeons also must a
void using mechanically stamped hardware that will be subject to cycli
c loads in vivo.