Delivery of radiofrequency (RF) energy from the distal tip of electrop
hysiology catheters produces lesions that may be too small to ablate a
rrhythmogenic sites during a single application of RF energy. To produ
ce larger lesions, we delivered RF energy via a quadripolar catheter i
n which all four electrodes were connected in unipolar fashion. The ca
theter (Webster Labs) had a 4-mm tip, 2-mm ring electrodes, and 2-mm i
nterelectrode distance. Lesion size was compared using RF energy deliv
ered in a multipolar configuration with that delivered only to the dis
tal tip using fresh bovine ventricular tissue. in vivo, RF lesions wer
e made in dogs using the distal tip as well as all four poles of the s
ame catheter inserted percutaneously. RF energy was delivered using a
constant voltage at a frequency of 400 kHz. Preliminary experiments we
re conducted to determine the maximum power deliverable without coagul
ation using each electrode configuration. The use of simultaneous mult
ipolar RF ablation produced significantly larger lesions both in vitro
and in vivo. The length of the lesion was increased by a factor of ap
proximately 2 in both the in vitro and in vivo experiments. There was
a trend toward an increasing depth of the lesion by simultaneously app
lying RF energy to all four electrodes. Lesion width was significantly
increased in the in vivo studies. We concluded that simultaneous mult
ipolar delivery of RF energy produces larger lesions than can be obtai
ned with delivery of RF energy to the distal tip alone. This technique
may offer a means of increasing lesion size, leading to a decrease in
the number of applications of RF energy necessary for ablation of arr
hythmias.