HPGE DETECTOR TRUE-COINCIDENCE CORRECTION FOR EXTENDED CYLINDER AND MARINELLI-BEAKER SOURCES

Citation
Tk. Wang et al., HPGE DETECTOR TRUE-COINCIDENCE CORRECTION FOR EXTENDED CYLINDER AND MARINELLI-BEAKER SOURCES, Nuclear instruments & methods in physics research. Section A, Accelerators, spectrometers, detectors and associated equipment, 376(2), 1996, pp. 192-202
Citations number
17
Categorie Soggetti
Nuclear Sciences & Tecnology","Physics, Particles & Fields","Instument & Instrumentation",Spectroscopy
ISSN journal
01689002
Volume
376
Issue
2
Year of publication
1996
Pages
192 - 202
Database
ISI
SICI code
0168-9002(1996)376:2<192:HDTCFE>2.0.ZU;2-#
Abstract
A semi-empirical method is developed to estimate the gamma-gamma true- coincidence correction factor (COI) for extended cylinder and Marinell i-beaker sources. Compared with point sources, the true-coincidence ef fects for extended sources are more complicated, since not only the so urce-volume averaged peak and total efficiencies but also the differen tial-efficiency distributions within the source itself are involved. A volume-effect factor (F-v). which can be calculated by the aid of our previously developed ESOLAN program, is introduced in this work to ha ndle these complexities. The validity of using this method was success fully demonstrated by comparing the predicted COI values with the expe rimentally determined values for several radionuclides.