Centrifugal pumps have become important devices for cardiopulmonary by
pass and circulatory assistance. Five types of centrifugal pumps are c
linically available in Japan. To evaluate the blood trauma caused by c
entrifugal pumps, a comparative hemolysis study was performed under id
entical conditions. In vitro hemolysis test circuits were constructed
to operate the BioMedicus BP-80 (Medtronic, BioMedicus), Sarns Delphin
(Sarns/ 3M Healthcare), Isoflow (St. Jude Medical [SJM]), HPM-15 (Nik
kiso), and Capiox CX-SP45 (Terumo). The hemolysis test loop consisted
of two 1.5 m lengths of polyvinyl chloride tubing with a 3/8-inch inte
rnal diameter, a reservoir with a sampling port, and a pump head. All
pumps were set to flow at 6 L/min against the total pressure head of 1
20 mm Hg. Experiments were conducted simulta neously for 6 h at room t
emperature (21 degrees C) with fresh bovine blood. Blood samples for p
lasma-free hemoglobin testing were taken, and the change in temperatur
e at the pump outlet port was measured during the experiment. The mean
pump rotational speeds were 1,570, 1,374, 1,438, 1,944, and 1,296 rpm
, and the normalized indexes of hemolysis were 0.00070, 0.00745, 0.000
96, 0.00066, 0.00090 g/100 L for the BP-80, Sarns, SJM, Nikkiso, and T
erumo pumps, respectively. The change in temperature at the pump outle
t port was the least for the Nikkiso pump (1.8 degrees C) and the most
with the SJM pump (3.8 degrees C). This study showed that there is no
relationship between the pump rotational speed (rpm) and the normaliz
ed index of hemolysis in 5 types of centrifugal pumps. The pump design
and number of impellers could be more notable factors in blood damage
.