B. Meyns et al., THE HEART-HEMOPUMP INTERACTION - A STUDY OF HEMOPUMP FLOW AS A FUNCTION OF CARDIAC ACTIVITY, Artificial organs, 20(6), 1996, pp. 641-649
The Hemopump is a useful left ventricular assist device. Because it is
a rotary blood pump, the pump performance is not constant and is depe
ndent on the cardiac cycle. We measured the static flow delivered by t
he pump at varying pressure heads (Delta P) in a mock circulation. The
se data are compared to the pump performance in vivo. On the basis of
these results, 5 sheep were instrumented for continuous Hemopump flow
measure ment as well as left ventricular and aortic pressure measureme
nts. The Hemopump flow was relayed instantaneously to the pressure hea
d. Low filling and ventricular failing (through intravenous administra
tion of a beta-blocker) conditions were applied. The in vivo measured
flows also are pressure head dependent, but the flow curve shows hyste
resis resulting in a loop during each cardiac cycle. The in vivo peak
flows (Delta P = 0) are similar to the in vitro data. The in vivo mean
s flows (a = 50 mm Hg) are similar to the in vitro data for the lower
pump speeds but are less than that at the higher pump speeds (3.74 +/-
0.55 L/min in vivo at Speed 7 versus 4.6 L/min in vitro). Low filling
interrupts the Delta P-flow loop and reduces flow. In the failing ven
tricle, Delta P increases and flow is reduced. The cannula leaks and r
esults in aortic insufficiency (0.36 +/- 0.05 L/min) when the pump is
turned off. Several conclusions have been drawn from these tests: Card
iac activity is beneficial for the pump performance as well as when th
e aortic pressure curve is nonpulsatile; the longer the systolic phase
, the higher the pump flow; the pump should never be turned off in cli
nical use; and filling is important for the pump's performance.